Two engineers measure the same oscillator 5 dB apart on the Signal Hound PN400, and "-5 dBm for higher SNR" made it worse — input level and mode conventions in phase-noise measurement
Symptom
A user measures a low-phase-noise crystal oscillator with the Signal Hound PN400. Engineer A uses the default setup and gets -142 dBc/Hz @ 10 kHz; engineer B sets up a different PC from scratch and gets -137 dBc/Hz, and both suspect one of the instruments is faulty. Later, the same crystal reads closer to A's value on a benchtop phase-noise analyzer. Someone else set the input power to -5 dBm "to get a better signal-to-noise ratio", and the whole trace degraded, with an abnormal rise appearing in the high-offset region.
Root-cause analysis
Phase-noise measurement is extremely sensitive to input level and operating-mode settings, and because the Signal Hound PN400 is a cross-correlation system built from two SMs, the settings must be identical across both instruments:
Input power outside the optimal region: the SM receiver channels have a recommended mixer drive level — too low and the channel's own noise enters the result (visible as an elevated far-offset floor, plus reduced correlation efficiency when the two channels disagree); too high and compression / AM-to-PM conversion makes "the DUT look noisier than it is". For many phase-noise measurement ranges, -5 dBm is far too hot: compression and distortion contributions mix into the reading;
Asymmetric input / attenuation / gain configuration between the two SMs: cross-correlation requires the two paths to be as homogeneous as possible. If one side is set to 0 dB attenuation and the other to 10 dB, the two channels have different floors and different linearity, and the correlated trace carries asymmetric artefacts;
Different measurement-mode conventions: the direct phase-noise method versus the discriminator method, and whether cross-correlation is enabled, each have their own effective floor and valid range — "default configuration" is not necessarily the same on freshly installed machines (different software versions, different templates). Most likely the 5 dB gap between A and B is a difference in mode and level, not in hardware;
Reference-level / power calibration differences: when the DUT's actual output power deviates from the set value, the effective drive point differs, producing dB-level drift between successive measurements.
Troubleshooting and standard practice
Fix a "measurement conditions table": for each DUT type, freeze the input power (start at the middle of the recommended region, e.g. around the -20 to -10 dBm class, optimised per the manual), the attenuation/gain settings of both SMs, the mode switches and the offset sweep table. Store them as a software template plus a text file, and have engineers A and B recall the same template for re-measurement;
Symmetry check: log each channel's live phase-noise trace separately — the two should roughly overlap. Clear separation means you check level/attenuation consistency between sides and single-channel health;
Find the optimal drive with a power sweep: run the DUT output through a step attenuator and sweep in 5–10 dB steps, plotting "reading versus drive power". The plateau in the middle is the linear measurement region; the two flanks are floor-limited and compression-limited respectively. Put the everyday operating point inside the plateau;
Align conventions with the benchtop instrument: before comparing against external equipment, unify offset points, level, temperature and supply conditions. If the difference exceeds 3 dB, re-check the template before blaming hardware;
Documentation: keep full screenshots of the measurement conditions for the standard reference source in the factory acceptance record, so every later deviation can first be checked against that table.
Acceptance criteria
Five consecutive runs on the reference source under the same template give repeatability ≤1 dB at each offset point; the two independent channel traces overlap well; the measured drive-power plateau has been archived and the operating point stated in the SOP.
(Compiled from the PN400 system architecture (dual-SM cross-correlation) and general practice for levels and modes in phase-noise measurement; the recommended drive region is subject to the manual shipped with the unit.)




Recently Posted
-
Cross‑correlation enabled on Signal Hound PN400; noise floor failed to drop 20 dB, trace grows noisier with averaging, cross‑correlation not convergin
November 26, 2026Signal Hound PN400 phase noise and VCO tester supplierI turned on cross-correlation in the Signal Hound PN400 but the floor didn
Read More -
Signal Hound PN400 supplier:Advertised floor -160 dBc/Hz, but my measurement flattens at -142
November 26, 2026Signal Hound PN400 phase noise and VCO tester supplierAdvertised floor -160 dBc/Hz, but my measurement flattens at -142 — what the
Read More -
Signal Hound PN400 Incoming Inspection Only Verified Power-Up; All Readings Drifted After Three Months,supplier
November 26, 2026Signal Hound PN400 phase noise and VCO tester supplierIncoming inspection only proved the Signal Hound PN400 "powers on"
Read More -
Signal Hound PN400 supplier: Moving Comb Spikes on Phase Noise Trace on Standard Lab Bench
November 26, 2026Signal Hound PN400 phase noise and VCO tester supplierWith the Signal Hound PN400 on an ordinary lab bench, the phase-noise trace
Read More