Signal Hound PN400 supplier:Advertised floor -160 dBc/Hz, but my measurement flattens at -142

Signal Hound PN400 phase noise and VCO tester supplier


Advertised floor -160 dBc/Hz, but my measurement flattens at -142 — what the Signal Hound PN400 residual noise floor is really made of, and how to calibrate it

Signal Hound PN400 supplier:Advertised floor -160 dBc/Hz, but my measurement flattens at -142


Symptom

Purchasing planned its test capability around the marketing page's "ultra-low residual noise floor". After delivery, measuring a crystal in the -155 dBc/Hz class gives a curve that goes flat near -142 dBc/Hz — suspicion that "the instrument isn't meeting its rating". Changing offset points and carrier frequencies only adds confusion: the floor is better at 10 kHz than at 1 kHz; a 5 GHz carrier sits several dB higher across the whole curve than a 1 GHz carrier; and after repositioning the two SMs the floor changes again.

Root-cause analysis

For a cross-correlation system like the Signal Hound PN400, the residual noise floor is not a constant line — it is a surface over "offset × carrier frequency", so comparing a single advertised number against an entire surface will always fail to match:

  1. The nominal figure equals the optimum point: the advertised residual (for example, a -168 dBc/Hz-class best region at low offsets for some configuration) is always obtained at a specific carrier frequency, a specific offset region, and specific level/bandwidth conditions. Move away from the optimum and the floor naturally rises — 1 kHz offset being worse than 10 kHz, and high carriers being worse than low carriers, are architectural regularities;

  2. Direct weighting from drive level and mode settings: if the mixer drive points on the two SMs deviate from the recommended region (see the level piece), the whole surface lifts by 2–5 dB. Measuring the system floor must first set the optimal link parameters per the manual — a floor obtained with casual settings is not the system's true capability;

  3. Channel mismatch and ageing: if the two SMs differ substantially in gain setting or port return loss, or one front end has degraded, the cancellation efficiency of the correlated common floor drops;

  4. System residual floor versus instrument noise figure: an SM's standalone DANL and "its contribution to the floor curve when used as a phase-noise channel" are two different things — you cannot derive the latter from the former;

  5. No terminated-floor baseline was taken at acceptance: without a record from factory or incoming inspection, later arguments about "how low should it actually be" have no basis.

Calibration and verification method

  1. Build a floor archive: using the manual's optimal link settings, measure the system residual at 5 / 10 / 20 / 40 (carrier, offset) combinations via two means — a reference artifact (low-phase-noise source) plus terminated-state checks — plot the "offset × carrier" residual-floor surface as a baseline, and store it with the equipment record;

  2. Compare against the manual / datasheet: check the baseline against the per-condition floor table in the shipped manual (a single marketing-page value serves only as an optimum-point reference). Landing inside the per-condition ranges means pass; only a uniform shift warrants suspecting link settings;

  3. Track degradation: re-measure the floor surface quarterly; if any region rises ≥3 dB, schedule a channel check (connector cleaning, level re-verification, single-unit self-calibration, and send-in for calibration if needed);

  4. Communicate the capability boundary (pre-sales / handover): promise customers the floor value read off the surface at their carrier frequency and offset region — more honest and more professional than quoting the lowest marketing figure. For DUTs below the floor, explain in advance the options of adding correlation counts or changing the paired model.

Acceptance criteria

The residual-floor surface baseline is documented and stored with the crate; every dispute about system capability can be settled against a comparison table of "per-condition measured values versus per-condition datasheet specifications"; the quarterly re-measurement plan is part of the maintenance regime.

(Compiled from Signal Hound's PN400 product material (dual-SM cross-correlation, per-configuration residual noise floor specifications) together with general practice for calibrating the floor of a phase-noise system; specific figures are subject to the paired models and the manual shipped with the unit.)